Gilgen-Ammann, Rahel; Buller, Mark J; Bitterle, Jacqueline L; Delves, Simon K; Veenstra, Bertil J; Roos, Lilian; Beeler, Nadja; Wyss, Thomas (2018). Evaluation of pulse rate measurement with a wrist worn device during different tasks and physical activity Current Issues in Sport Science, 3, pp. 1-9. Innsbruck University Press 10.15203/CISS_2018.011
|
Text
Ammann-Gilgen_2018_Evaluation of pulse rate measurement with a wrist worn device during different tasks and physical activity.pdf - Published Version Available under License Creative Commons: Attribution-Noncommercial (CC-BY-NC). Download (597kB) | Preview |
The purpose of this study was to evaluate the wrist-worn device Mio FUSE, which estimates heart rate (HR) based on photo-plethysmography, 1) in a large study group during a standardised activity, 2) in a small group during a variety of activities and 3) to investigate factors affecting HR accuracy in a real-world setting. First, 53 male participants (20 ±1 years; 1.79 ±0.07 m; 76.1 ±10.5 kg) completed a 35-km march wearing the Equivital EQ-02 as a criterion measure. Second, 5 participants (whereof 3 female; 29 ±5 years; 1.74 ±0.07 m; 67.8 ±11.1 kg) independently performed 25 activities, categorised as sitting passive, sitting active, standing, cyclic and anti-cyclic activities with the Polar H7 as a criterion device. Equivalence testing and Bland-and-Altman analyses were undertaken to assess the accuracy to the criterion devices. Third, confounders affecting HR accuracy were investigated using multiple backwards regression analyses. The Mio FUSE was equivalent to the respective criterion measures with only small systematic biases of -3.5 bpm (-2.6%) and -1.7 bpm (-1.3%) with limits of agreements of ±10.1 bpm and ±10.8 bpm during the 35-km march and during different activities, respectively. Confounding factors negatively affecting the accuracy of the Mio FUSE were found to include larger wrist size and intensified arm and/or wrist movement. The wrist-worn Mio FUSE can be recommended to estimate overall HR accurately for different types of activities in healthy adults. However, during sporting activities involving intensified arm and/or wrist movement or for detailed continuous analysis, a chest strap is preferred to the Mio FUSE to optimise HR estimation accuracy.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
Swiss Federal Institute of Sports Magglingen SFISM > EHSM - Lehre und Sportpädagogik > Monitoring |
Name: |
Gilgen-Ammann, Rahel; Buller, Mark J; Bitterle, Jacqueline L; Delves, Simon K; Veenstra, Bertil J; Roos, Lilian; Beeler, Nadja and Wyss, Thomas |
ISSN: |
2414-6641 |
Publisher: |
Innsbruck University Press |
Language: |
English |
Submitter: |
Service Account |
Date Deposited: |
03 May 2021 09:38 |
Last Modified: |
03 Oct 2021 02:18 |
Publisher DOI: |
10.15203/CISS_2018.011 |
Related URLs: |
|
Uncontrolled Keywords: |
Fitness tracker Photo-plethysmography Measurement accuracy |
ARBOR DOI: |
10.24451/arbor.11001 |
URI: |
https://arbor.bfh.ch/id/eprint/11001 |