Repository logo
  • English
  • Deutsch
  • Français
Log In
New user? Click here to register.Have you forgotten your password?
  1. Home
  2. CRIS
  3. Publication
  4. RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise.
 

RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise.

URI
https://arbor.bfh.ch/handle/arbor/40583
Version
Published
Date Issued
2019
Author(s)
Gilgen-Ammann, Rahel
Schweizer, Theresa
Wyss, Thomas  
Type
Article
Language
English
Subjects

Gold standard Heart r...

Abstract
This study was designed to examine the RR interval signal qualities of a Holter device and a heart rate chest belt monitor at rest and during exercise. Ten healthy individuals completed five low- to high-intensity activities while simultaneously using the medilog; ®; AR12plus? Holter monitor and the Polar H10 heart rate monitor. The RR interval signal quality was based on the quantification of the missing RR intervals and RR interval detection errors. Therefore, both measurement systems were compared against visual inspection of the raw electrocardiography signal. The missing and wrong R-wave peak detections were counted manually for both measurement systems. RR interval signal quality was defined as the relative number of correctly detected RR intervals. Overall, RR interval signal qualities of 94.6% and 99.6% were demonstrated for the medilog® AR12plus and the Polar H10. During the high-intensity activities, the RR interval signal quality of the medilog® AR12plus dropped to 89.8%, whereas the Polar H10 maintained a signal quality of 99.4%. The correlation between both systems was high (r=0.997, p>0.001). The excellent RR interval signal quality during low- to moderate-intensity activities in the medilog® AR12plus and during low- to high-intensity activities in the Polar H10 demonstrates both measurement systems' validity for the detection of RR intervals throughout a wide range of activities. A simple chest strap such as the Polar H10 might be recommended as the gold standard for RR interval assessments if intense activities with strong body movements are investigated.
DOI
10.24451/arbor.11053
https://doi.org/10.24451/arbor.11053
Publisher DOI
10.1007/s00421-019-04142-5
Journal or Serie
European Journal of Applied Physiology
ISSN
1439-6327
Publisher URL
https://link.springer.com/article/10.1007/s00421-019-04142-5
Related URL
https://link.springer.com/content/pdf/10.1007/s00421-019-04142-5.pdf publication
Organization
Eidgenössische Hochschule für Sport Magglingen (nur "virtuell" für ARBOR)  
Monitoring  
Volume
119
Issue
7
Publisher
Springer
Submitter
ServiceAccount
Citation apa
Gilgen-Ammann, R., Schweizer, T., & Wyss, T. (2019). RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise. In European Journal of Applied Physiology (Vol. 119, Issue 7). Springer. https://doi.org/10.24451/arbor.11053
File(s)
Loading...
Thumbnail Image

restricted

Name

GilgenAmmann_RRIntervalSignalQuality_2019.pdf

License
Publisher
Version
published
Size

721.79 KB

Format

Adobe PDF

Checksum (MD5)

5655f447ff00d8c47330cd2989306c61

About ARBOR

Built with DSpace-CRIS software - System hosted and mantained by 4Science

  • Cookie settings
  • Privacy policy
  • End User Agreement
  • Send Feedback
  • Our institution