Mechanical Design and Control System Development of a Rehabilitation Robotic System for Walking With Arm Swing
Version
Published
Date Issued
2021-11-18
Author(s)
Type
Article
Language
English
Abstract
Background: Interlimb neural coupling implies that arm swing should be included during gait training to improve rehabilitation outcomes. We previously developed several systems for production of walking with arm swing, but the reaction forces on the foot sole during usage of the systems were not satisfactory and there was potential to improve control system performance. This work aimed to design and technically evaluate a novel system for producing walking with synchronised arm and leg movement and with dynamic force loading on the foot soles.
Methods: The robotic system included a passive curved treadmill and a trunk frame, upon which the rigs for the upper and lower limbs were mounted. Ten actuators and servocontrollers with EtherCAT communication protocol controlled the bilateral shoulder, elbow, hip, knee and ankle joints. Impedance control algorithms were developed and ran in an industrial PC. Flexible pressure sensors recorded the plantar forces on the foot soles. The criteria of implementation and responsiveness were used to formally evaluate the technical feasibility of the system.
Results: Using impedance algorithms, the system produced synchronous walking with arm swing on the curved treadmill, with mean RMS angular tracking error <2° in the 10 joint profiles. The foot trajectories relative to the hip presented similar shapes to those during normal gait, with mean RMS displacement error <1.5 cm. A force pattern that started at the heel and finished at the forefoot was observed during walking using the system, which was similar to the pattern from overground walking.
Conclusion: The robotic system produced walking-like kinematics in the 10 joints and in the foot trajectories. Integrated with the curved treadmill, the system also produced walking-like force patterns on the foot soles. The system is considered feasible as far as implementation and responsiveness are concerned. Future work will focus on improvement of the mechanical system for future clinical application.
Methods: The robotic system included a passive curved treadmill and a trunk frame, upon which the rigs for the upper and lower limbs were mounted. Ten actuators and servocontrollers with EtherCAT communication protocol controlled the bilateral shoulder, elbow, hip, knee and ankle joints. Impedance control algorithms were developed and ran in an industrial PC. Flexible pressure sensors recorded the plantar forces on the foot soles. The criteria of implementation and responsiveness were used to formally evaluate the technical feasibility of the system.
Results: Using impedance algorithms, the system produced synchronous walking with arm swing on the curved treadmill, with mean RMS angular tracking error <2° in the 10 joint profiles. The foot trajectories relative to the hip presented similar shapes to those during normal gait, with mean RMS displacement error <1.5 cm. A force pattern that started at the heel and finished at the forefoot was observed during walking using the system, which was similar to the pattern from overground walking.
Conclusion: The robotic system produced walking-like kinematics in the 10 joints and in the foot trajectories. Integrated with the curved treadmill, the system also produced walking-like force patterns on the foot soles. The system is considered feasible as far as implementation and responsiveness are concerned. Future work will focus on improvement of the mechanical system for future clinical application.
Publisher DOI
Journal
Frontiers in Rehabilitation Sciences
ISSN
2673-6861
Volume
2
Publisher
Frontiers Media SA
Submitter
Fang, Juan
Citation apa
Fang, J., & Hunt, K. J. (2021). Mechanical Design and Control System Development of a Rehabilitation Robotic System for Walking With Arm Swing. In Frontiers in Rehabilitation Sciences (Vol. 2). Frontiers Media SA. https://doi.org/10.24451/arbor.8761
File(s)![Thumbnail Image]()
Loading...
open access
Name
fresc-02-720182.pdf
License
Attribution 4.0 International
Version
published
Size
2.82 MB
Format
Adobe PDF
Checksum (MD5)
6e234461ba7a0a52f86137df0a67c2d4