Repository logo
  • English
  • Deutsch
  • Français
Log In
New user? Click here to register.Have you forgotten your password?
  1. Home
  2. CRIS
  3. Publication
  4. Robust control of heart rate for cycle ergometer exercise
 

Robust control of heart rate for cycle ergometer exercise

URI
https://arbor.bfh.ch/handle/arbor/40698
Version
Published
Date Issued
2019
Author(s)
Hunt, Kenneth James  
Hurni, Cédric C.
Type
Article
Language
English
Abstract
The objective was to assess the performance and robustness of a novel strategy for automatic control of heart rate (HR) during cycle ergometry. Control design used a linear plant model and direct shaping of the closed-loop input-sensitivity function to achieve an appropriate response to disturbances attributable to broad-spectrum heart rate variability (HRV). The controller was evaluated in 73 feedback control experiments involving 49 participants. Performance and stability robustness were analysed using a separately identified family of 73 plant models. The controller gave highly accurate and stable HR tracking performance with mean root-mean-square tracking error between 2.5 beats/min (bpm) and 3.1 bpm, and with low average control signal power. Although plant parameters varied over a very wide range, key closed-loop transfer functions remained invariant to plant uncertainty in important frequency bands, while infinite gain margins and large phase margins (>62◦) were preserved across the whole plant model family. Highly accurate, stable and robust HR control can be achieved using LTI controllers of remarkably simple structure. The results highlight that HR control design must focus on disturbances caused by HRV. The input-sensitivity approach evaluated in this work provides a transparent method of addressing this challenge.
Subjects
TA Engineering (General). Civil engineering (General)
DOI
10.24451/arbor.8708
https://doi.org/10.24451/arbor.8708
Publisher DOI
10.1007/s11517-019-02034-6
Journal or Serie
Medical & Biological Engineering & Computing
ISSN
1741-0444
Publisher URL
https://doi.org/10.1007/s11517-019-02034-6
Organization
Institut für Rehabilitation und Leistungstechnologie IRPT  
BFH-Zentrum Technologien in Sport und Medizin  
Digital Health Lab  
Technik und Informatik  
BFH-Zentren  
Institut für Medizininformatik I4MI  
Volume
57
Issue
11
Publisher
Springer
Submitter
HuntK
Citation apa
Hunt, K. J., & Hurni, C. C. (2019). Robust control of heart rate for cycle ergometer exercise. In Medical & Biological Engineering & Computing (Vol. 57, Issue 11, pp. 2471–2482). Springer. https://doi.org/10.24451/arbor.8708
File(s)
Loading...
Thumbnail Image
Download

open access

Name

HunHur19.pdf

License
Attribution 4.0 International
Version
published
Size

1.74 MB

Format

Adobe PDF

Checksum (MD5)

12b61b91cc105376d4e802d5a593e797

About ARBOR

Built with DSpace-CRIS software - System hosted and mantained by 4Science

  • Cookie settings
  • Privacy policy
  • End User Agreement
  • Send Feedback
  • Our institution