Repository logo
  • English
  • Deutsch
  • Français
Log In
New user? Click here to register.Have you forgotten your password?
  1. Home
  2. CRIS
  3. Publication
  4. Development and evaluation of a titanium-based planar ultrasonic scalpel for precision surgery
 

Development and evaluation of a titanium-based planar ultrasonic scalpel for precision surgery

URI
https://arbor.bfh.ch/handle/arbor/35947
Version
Published
Date Issued
2023
Author(s)
Hofmann, Martin  
Haeberlin, Andreas
de Brot, Simone
Stahel, Andreas  
Keppner, Herbert
Burger, Jürgen  
Type
Article
Language
English
Subjects

Ultrasonic scalpel Ul...

Abstract
This paper introduces a titanium-based planar ultrasonic microscalpel. The concept of silicon-based planar ul­trasonic transducers has already been proven, but they are not yet suitable for clinical use due to material failure. The main objective of this work was to develop a smaller, lighter, and more cost-effective ultrasonic scalpel that could be used as an alternative or supplementary device to current surgical instruments. Various prototypes were fabricated and characterized, differing in bonding by three epoxy adhesives and two solder pastes as well as three variations in tip design. The instruments were designed to operate in the frequency range of commercial in­ struments and to generate a longitudinal displacement amplitude. The electro-mechanical characterization through impedance analysis and vibration measurements was complemented by an in vitro cutting trial and an acute in vivo animal experiment in comparison to commercial ultrasonic and electrosurgical devices. The operating frequency was around 40 kHz and 48 kHz depending on whether matched or unmatched operation was used. Unmatched operation turned out to be more suitable, achieving displacement amplitudes of 25.3 μm and associated velocity amplitudes of up to 7.9 m/s at an electrical power of 10.2 W. The cutting ability was demonstrated in vivo by successful dissection even under anticoagulation. The geometry of the instrument tip was found to have a major influence on cutting performance by affecting the resonance behaviour and tissue penetration.
Subjects
QC Physics
DOI
10.24451/arbor.18774
https://doi.org/10.24451/arbor.18774
Publisher DOI
10.1016/j.ultras.2023.106927
Journal
Ultrasonics
ISSN
0041-624X
Publisher URL
https://www.sciencedirect.com/science/article/pii/S0041624X23000033
Organization
HUCE / Labor für Sensorik  
Technik und Informatk  
Institute for Human Centered Engineering (HUCE)  
Volume
130
Publisher
Elsevier
Submitter
StahelA
Citation apa
Hofmann, M., Haeberlin, A., de Brot, S., Stahel, A., Keppner, H., & Burger, J. (2023). Development and evaluation of a titanium-based planar ultrasonic scalpel for precision surgery. In Ultrasonics (Vol. 130). Elsevier. https://doi.org/10.24451/arbor.18774
File(s)
Loading...
Thumbnail Image

open access

Name

1-s2.0-S0041624X23000033-main.pdf

License
Attribution 4.0 International
Version
published
Size

13.89 MB

Format

Adobe PDF

Checksum (MD5)

8a42b83fd24f125e8b82cc63438e41f0

About ARBOR

Built with DSpace-CRIS software - System hosted and mantained by 4Science

  • Cookie settings
  • Privacy policy
  • End User Agreement
  • Send Feedback
  • Our institution